000988379 001__ 988379 000988379 005__ 20251028135621.0 000988379 0247_ $$2HBZ$$aHT030782669 000988379 0247_ $$2Laufende Nummer$$a43436 000988379 0247_ $$2datacite_doi$$a10.18154/RWTH-2024-06145 000988379 037__ $$aRWTH-2024-06145 000988379 041__ $$aEnglish 000988379 082__ $$a620 000988379 1001_ $$0P:(DE-82)IDM05990$$aEhreiser, Sonja$$b0$$urwth 000988379 245__ $$aOptimization of patient matching in total knee arthroplasty through improved implant sizing and detailed fit assessment$$cvorgelegt von Sonja Anke Gabriele Agathe Ehreiser (geb. Grothues)$$honline 000988379 246_3 $$aOptimierung der Patienten-Anpassung in der totalen Knieendoprotethik durch verbesserte Implantat-Größen-Verteilung und detaillierte Fit Bewertung$$yGerman 000988379 260__ $$aAachen$$bRWTH Aachen University$$c2024 000988379 300__ $$a1 Online-Ressource : Illustrationen 000988379 3367_ $$02$$2EndNote$$aThesis 000988379 3367_ $$0PUB:(DE-HGF)11$$2PUB:(DE-HGF)$$aDissertation / PhD Thesis$$bphd$$mphd 000988379 3367_ $$2BibTeX$$aPHDTHESIS 000988379 3367_ $$2DRIVER$$adoctoralThesis 000988379 3367_ $$2DataCite$$aOutput Types/Dissertation 000988379 3367_ $$2ORCID$$aDISSERTATION 000988379 500__ $$aVeröffentlicht auf dem Publikationsserver der RWTH Aachen University 000988379 502__ $$aDissertation, Rheinisch-Westfälische Technische Hochschule Aachen, 2024$$bDissertation$$cRheinisch-Westfälische Technische Hochschule Aachen$$d2024$$gFak04$$o2024-05-29 000988379 5203_ $$aDer totale Kniegelenkersatz ist eine der meist-durchgeführten Operationen weltweit. Indikatoren des Operationserfolgs wie Revisionsraten oder die Patientenzufriedenheit sind multifaktoriell beeinflusst, wobei das Implantat Design und die zugehörige Passgenauigkeit einen relevanten und veränderlichen Faktor darstellt. Für diverse Patienten Populationen wurde eine erhebliche Diskrepanz zwischen Implantat- und Knochen-Form berichtet, hinsichtlich Größen sowie einzelner Morphologie Parameter. Solch geringe Passgenauigkeit kann die Empfehlung eines Individual-Implantats motivieren. Bislang gibt es jedoch weder standardisierte Methoden welche diese Entscheidung unterstützen, noch generell für eine detaillierte Passform-Bewertung. Das Ziel dieser Arbeit ist demzufolge die Potenzial-Bewertung einer Implantat-Größen Optimierung sowie die Entwicklung von Methoden zur individuellen Passformbewertung, für eine insgesamt optimierte Patienten-Anpassung. Die Möglichkeit einer verbesserten Patienten-Anpassung durch numerisch optimierte Implantatgrößen wurde für eine Datenbank von über 85.000 Knien demonstriert. Mit den optimierten Größen wurde eine um 19% bis 26% verbesserte Patientenabdeckung erreicht, im Vergleich zu bestehenden Systemen. Durch die exemplarische Dokumentation größen-unabhängiger Formvariation der femoralen J-Curve sowie relevanter Formabweichungen trotz idealem Größen-Fit, wurde der Bedarf an weiterer Passformanalyse gezeigt. In einer Literaturrecherche wurden Kriterien für die Passformbewertung identifiziert und als Interface, Morphologie, Alignment und Andere klassifiziert. Da die identifizierten, morphologischen Kriterien für das physiologische, nicht-deformierte Knie definiert sind, wurde eine parameter-basierte Deformitätskorrektur entwickelt und verifiziert. Basierend auf der Analyse bestehender Bildgebungsverfahren wurden Konzepte zur Fit-Bewertung definiert. Das gewählte Konzept beinhaltet eine initiale Risikoabschätzung mit standardisierten Röntgenbildern und, wenn erforderlich, zusätzlicher Detailanalyse mittels Computer-Tomographie. In einer beispielhaften Implementierung mit hoher Automatisierung wurde das Potenzial für die klinische Integration demonstriert. Die Passformbewertung wurde für ein beispielhaftes Implantat und eine große Kniedatenbank durchgeführt. Mit optimierten Implantatgrößen und individueller Passformbewertung kann der Anteil adäquat versorgter Patienten erhöht und diejenigen, welche ein Individual-Implantat benötigen, objektiv identifiziert werden, was insgesamt die Patienten-Anpassung maximiert.$$lger 000988379 520__ $$aTotal knee arthroplasty is one of the most performed surgeries worldwide. Indicators of treatment success, such as revision rates or patient satisfaction, are subject to multiple influences, with implant design and associated patient-specific implant fit being a relevant and variable factor. A considerable discrepancy between implant and bone shape in terms of size as well as individual morphologic parameters has been reported for various populations. Such mismatch can motivate the advising of a customized implant. However, there is currently no standardized method to support this decision, or for detailed assessment of knee implant fit in general. Hence, the aim of this thesis was to evaluate the potential for standard implant size optimization and to develop methods for patient-specific implant fit evaluation, with the goal of optimizing patient matching in total knee arthroplasty. The possibility of improved size matching through numerically optimized sizing was demonstrated for a large database of over 85.000 knees. With the numerically optimized sizes an increase in population coverage of 19% to 26% compared to representative existing implant systems was reached. The need for further fit analysis beyond size was demonstrated by an exemplary documentation of size-independent shape variations using the femoral J-curve as an example, and by relevant morphological deviations even in the case of an ideal size fit. Many objective criteria for assessing implant fit have been defined in the literature, which were assigned to the categories interface, morphological, alignment and other. As morphological criteria are defined for the physiological, non-deformed knee morphology, a parameter-based deformity correction was developed and verified. Based on the analysis of existing imaging procedures, concepts for the evaluation of the identified criteria were defined. The chosen concept includes an initial fit assessment based on standardized radiographs and, if required, an additional detailed fit evaluation using computed tomography data. The potential for clinical integration was demonstrated in an exemplary implementation with a high level of automation. The fit evaluation was carried out for an exemplary implant and a large knee database, demonstrating the workflows robustness. With the implant size optimization and individual fit evaluation, more patients can be provided with an adequate standard implant and those requiring a customized implant identified objectively, thereby maximizing patient matching.$$leng 000988379 588__ $$aDataset connected to Lobid/HBZ 000988379 591__ $$aGermany 000988379 653_7 $$aImplantat-Größen-Optimierung 000988379 653_7 $$aIndividual-Implantat 000988379 653_7 $$aKniemorphologie 000988379 653_7 $$aPassform-Bewertung 000988379 653_7 $$aimplant mismatch 000988379 653_7 $$aknee morphology 000988379 653_7 $$apatient-specific fit evaluation 000988379 653_7 $$asize optimization 000988379 7001_ $$0P:(DE-82)IDM00603$$aRadermacher, Klaus$$b1$$eThesis advisor$$urwth 000988379 7001_ $$0P:(DE-82)991095$$aStindel, Eric$$b2$$eThesis advisor 000988379 8564_ $$uhttps://publications.rwth-aachen.de/record/988379/files/988379.pdf$$yOpenAccess 000988379 8564_ $$uhttps://publications.rwth-aachen.de/record/988379/files/988379_source.docx$$yRestricted 000988379 909CO $$ooai:publications.rwth-aachen.de:988379$$popenaire$$popen_access$$pVDB$$pdriver$$pdnbdelivery 000988379 9101_ $$0I:(DE-588b)36225-6$$6P:(DE-82)IDM05990$$aRWTH Aachen$$b0$$kRWTH 000988379 9101_ $$0I:(DE-588b)36225-6$$6P:(DE-82)IDM00603$$aRWTH Aachen$$b1$$kRWTH 000988379 9141_ $$y2024 000988379 915__ $$0StatID:(DE-HGF)0510$$2StatID$$aOpenAccess 000988379 9201_ $$0I:(DE-82)419410_20140620$$k419410$$lLehrstuhl für Medizintechnik$$x0 000988379 961__ $$c2024-08-16T15:52:56.268085$$x2024-06-27T13:08:21.419641$$z2024-08-16T15:52:56.268085 000988379 9801_ $$aFullTexts 000988379 980__ $$aI:(DE-82)419410_20140620 000988379 980__ $$aUNRESTRICTED 000988379 980__ $$aVDB 000988379 980__ $$aphd